The CM4375-50 is a 1000 A AC/DC True RMS clamp meter that features a uniquely designed slim jaw that lets you easily get into the narrow gaps between cables in crowded electrical distribution boxes without fear of damaging the cables or the clamp meter. The broad IP54 dustproof and waterproof enclosure to enable professionals to work quickly in challenging situations including the development, repair, and maintenance of equipment and devices such as electric vehicles.
- DC High V Probe Mode enables high voltage measurement up to CAT III 2000 V
- When the Wireless Adapter Z3210 is installed, Bluetooth wireless communication is enabled
- When combined with GENNECT Cross, you can perform simple harmonic analysis
- Convenient for repeated measurement and recording tasks with automatic hold function
- Assess transient currents at motor startup
DC Current range | 1000 A, (Max. display 999.9 A) , Basic accuracy: ±1.3% rdg. ±0.3 A (at 30.1 A - 999.9 A) |
---|---|
AC Current range | 1000 A (Max. display 999.9 A, 10 Hz to 1 kHz, True RMS), Basic accuracy 45-66 Hz: ±1.8% rdg. ±0.3 A (at 30.1 A - 900.0 A) |
Crest factor | 1000 A range: 1.5 |
DC+AC Current range | 1000 A (DC, 10 Hz to 1 kHz, True RMS), Basic accuracy DC, 45-66 Hz: ±1.3% rdg. ±1.3 A (at 30.1 A - 900.0 A) |
DC Power range | 0.000 kVA to 1000 kVA (When using P2000: 0 kVA to 2000 kVA) (Automatically switched based on voltage range), Basic accuracy: ±2.0% rdg. ±20 dgt. |
DC Voltage range | 600.0 mV to 1000 V (When using P2000: 600.0 V to 2000 V) |
AC Voltage range | 6.000 V to 1000 V, 4 ranges (15 Hz to 1 kHz, True RMS), Basic accuracy 45 - 66 Hz: ±0.9% rdg. ±0.003 V (at 6 V) |
DC+AC Voltage range | 6.000 V to 1000 V, 4 ranges, Basic accuracy DC, 45-66 Hz: ±1.0% rdg. ±0.013 V (at 6 V) |
Resistance range | 600.0 Ω to 6.000 MΩ, 5 ranges, Basic accuracy: ±0.7% rdg. ±0.5 Ω (at 600 Ω) |
Electrostatic capacity range | 1.000 μF to 1000 μF, 4 ranges, Basic accuracy: ±1.9% rdg. ±0.005 μF (at 1 μF) |
Frequency range | 9.999 Hz to 999.9 Hz, 3 ranges, Basic accuracy: ±0.1% rdg. ±0.003 Hz (at 9.999 Hz) |
Temperature (K) | -40.0°C to 400.0°C, add temperature probe accuracy to basic accuracy of ±0.5% rdg. ±3.0°C |
Harmonic levels | [With Z3210 installed] Voltage/Current harmonic levels up to 30th, Content factor, Total harmonic distortion ratio *Harmonics can be displayed with our free app GENNECT Cross. |
Other functions | Continuity check, Diode check, Automatic AC/ DC detection, DC current and DC voltage polarity detection function, MAX/MIN/AVG/ PEAK MAX/PEAK MIN value display, Low-pass filter function, Display value hold, Auto hold, Backlight, Auto power save, Buzzer sound, Zero-adjustment |
Dustproof, waterproof | IP20 (Voltage measurement in a completely dry condition. When jaw closes) IP54 (While in storage) |
Power supply | LR03 Alkaline battery ×2 Continuous use: approx. 40 hr (without Z3210 installed), approx. 20 hr. (with Z3210 installed and using wireless communications) Other conditions: 100 A AC measurement, backlight off, 23°C reference value |
Core jaw diameter | φ34 mm (1.34 in) |
Smallest dimension of jaw cross-section | 9.5 mm (0.37 in) (Range value of 44 mm (1.73 in) from the tip of the jaw) |
Dimensions and mass | 65 mm (2.56 in) W × 242 mm (9.53 in) H × 35 mm (1.38 in) D mm, 350 g (12.3 oz) |
Included accessories | Test Lead L9300, Carrying Case C0203, LR03 Alkaline battery ×2, Instruction Manual ×2, Operating Precautions ×1 |